Determination of moisture/Water content by Karl Fischer Method
Determination of moisture/Water content by Karl Fischer Method
Equipment : Karl Fischer Titrator
Reagents : 1. AR grade Methanol
2. Pyridine free Karl Fischer Reagent
Procedure:
KF Factor Determination: Put methanol in the titration beaker enough to immerse the tip of platinum electrode.
Start stirrer and press start switch
Neutralize the methanol by titrating with KF-reagent.
Red indicating bulb glows at the end point.
Add one drop of water from the dropper of a previously weighed vial containing distilled water and reweigh to note the weight of water added.
Start titration by pressing the green start switch.
Note the volume of KF reagent required to neutralized the water (V)
KF Factor = W = mg/ml
V
Where:
W= Weight of water added.
V= Volume of KF reagent used.
0 Comments: